Remote Control
R&S
®
RTM20xx
303
User Manual 1317.4726.02 ─ 01
Usage:
Event
TRIGger:A:SOURce
<Source>
Sets the trigger source for the selected A trigger type.
Parameters:
<Source>
CH1 | CH2 | CH3 | CH4 | EXTernanalog | LINE | SBUS1 .. SBUS4 |
D0..D15
CH1 | CH2 | CH3 | CH4
One of the input channels is the trigger source. Available channels
depend on the instrument type.
EXTernanalog
External Trigger Input on the rear panel
LINE
AC line for the edge trigger
SBUS
Serial buses 1 to 4
D0..D15
If MSO option R&S
RTM-B1 is installed, the digital channels D0 to
D15 can be used as trigger sources for edge, width and pattern
trigger.
TRIGger:EXTern:COUPling
<ExternCoupling>
Sets the coupling for the external trigger input. The command is relevant if
is set to
EXTernanalog
.
Parameters:
<ExternCoupling>
AC | DC
*RST:
AC
TRIGger:A:TYPE
<Type>
Sets the trigger type for the A trigger.
Parameters:
<Type>
EDGE | WIDTh | TV | BUS | LOGic
EDGE: edge trigger
WIDTh: width trigger
TV: video trigger
BUS: only available if at least one option R&S RTM-K1 or R&S
RTM-K2 is installed.
LOGic: pattern trigger, logic trigger
See:
chapter 11, "Protocol Analysis"
Command Reference