Rise or falling time of the left-most rising edge within the displayed
section of the waveform of the selected channel. The reference
level for this mesurement is set with
PDCycle | NDCycle
Measure the positive or negative duty cycle.
PPWidth | NPWidth
Measure the width of positive or negative pulses.
CYCMean
Mean value of the left-most signal period of the waveform of the
selected channel.
CYCRms
RMS (Root Mean Square) value of the voltage of the left-most
signal period of the waveform of the selected channel.
STDDev
Measures the standard deviation of the waveform.
CYCStddev
Measures the standard deviation of one cycle, usually of the first,
left-most signal period.
TFRequency | TPERiode
Measure the frequency of the trigger signal and the length of the
its periods (hardware counter).
DELay
Time difference between two edges of the same or different wave-
forms. The waveforms are selected with
.
PHASe
Phase difference between two waveforms (time difference/period
* 360). The waveforms are selected with
BWIDth
Burst width, the duration of one burst, measured from the first edge
to the last edge that cross the middle reference level.
POVershoot | NOVershoot
Positive and negative overshoot of a square wave.
TBFRequency | TBPeriod
Measures the frequency of the B-trigger signal and the length of
the B-trigger signal periods.
*RST:
NONE (measurement is off)
MEASurement<m>:SOURce
<SignalSource>[,<ReferenceSource>]
Selects one of the active signal, reference or math channels as the source(s) of the
selected measurement. Available sources depend on the selected measurement type.
Command Reference