Remote Control
R&S
®
RTM20xx
363
User Manual 1317.4726.02 ─ 01
Parameters:
<Polarity>
POSitive | NEGative
*RST:
POS
Example:
chapter 15.3.2.1, "Searching for a Pulse of Specified Width"
Firmware/Software:
FW 03.400
SEARch:TRIGger:WIDTh:LEVel
<Level>
Sets the voltage level on which the pulse width is measured.
Parameters:
<Level>
*RST:
500 mV
Firmware/Software:
FW 03.400
SEARch:TRIGger:WIDTh:LEVel:DELTa
<DeltaLevel>
Sets a hysteresis range above and below the search level to avoid unwanted search
results caused by noise oscillation around the level.
Parameters:
<DeltaLevel>
Range:
Lower limit depends on vertical scale and other set-
tings, no upper limit
*RST:
200 mV
Firmware/Software:
FW 03.400
SEARch:TRIGger:WIDTh:RANGe
<Range>
Sets how the measured pulse width is compared with the given limit(s).
To set the width, use
To set the range ± Δt, use
Parameters:
<Range>
WITHin | OUTSide | SHORter | LONGer
WITHin
Finds pulses inside the range
width ± Δt
.
OUTSide
Finds pulses outside the range
width ± Δt
.
SHORter
Finds pulses shorter than the given width.
LONGer
Finds pulses longer than the given width.
*RST:
WITH
Example:
chapter 15.3.2.1, "Searching for a Pulse of Specified Width"
Command Reference