Remote Control
R&S
®
RTM20xx
297
User Manual 1317.4726.02 ─ 01
........................................................................................300
.................................................................................................301
............................................................................................301
............................................................................................301
PROBe<m>:SETup:ATTenuation[:AUTO]?
Returns the attenuation of an automatically detected probe.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Return values:
<ProbeAttenuation> Range:
1 to 1000
Usage:
Query only
PROBe<m>:SETup:ATTenuation:UNIT
<Unit>
Selects the unit that the probe can measure.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Unit>
V | A
Firmware/Software:
FW 03.700
PROBe<m>:SETup:ATTenuation:MANual
<ManualAttenuation>
Sets the attenuation or gain of the probe if the probe was not detected by the instrument.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<ManualAttenuation> Range:
1 to 10000
*RST:
1
Firmware/Software:
FW 03.700
PROBe<m>:SETup:BANDwidth?
Queries the bandwidth of the probe.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Command Reference