![Rohde & Schwarz EVSG-B1 User Manual Download Page 154](http://html.mh-extra.com/html/rohde-and-schwarz/evsg-b1/evsg-b1_user-manual_1477820154.webp)
Graphic Measurement Modes
R&S
®
EVSG1000
154
User Manual 1178.6227.02 ─ 06
Manual mode
Select a predefined fixed value manually.
Remote command:
IF BW
Defines the IF demodulation bandwidth for RX input in "AF Spectrum" mode.
Remote command:
AF In BW
Access
: [BW]
Access VNC
: [e]
Bandwidth with which the LF input is measured.
Remote command:
IF BW AM
Defines the IF demodulation bandwidth for the AM component of LF input.
This function is only available for AM carrier input, which cannot be selected directly in
"AF Spectrum" mode. However, it becomes available when you do one of the following:
●
Switch to the AF spectrum from "ILS LOC" or "ILS GP" mode in low IF mode
●
Switch to the AF spectrum from "VOR" mode for LF input, selecting the AM
demodulated subcarrier (see
"Switching from numeric to graphical results"
Remote command:
5.3.4
Configuring the Amplitude (Y-Axis) for RF Input
Access
: [Ampt]
Access VNC
: [w]
The amplitude settings configure the y-axis of the spectrum. For the AF Spectrum, the
modulation depth of the AM signal components is displayed on the y-axis.
The following settings are available for RF input.
Y-Range
Determines the displayed modulation depth range on the y-axis of the spectrum.
Remote command:
AF Spectrum Mode (Option R&S
EVSG-K11)