![Rohde & Schwarz EVSG-B1 User Manual Download Page 86](http://html.mh-extra.com/html/rohde-and-schwarz/evsg-b1/evsg-b1_user-manual_1477820086.webp)
Numeric Measurement Modes
R&S
®
EVSG1000
86
User Manual 1178.6227.02 ─ 06
"Narrow"
Narrow filter; the DDM filters used to determine the 90
Hz/ 150
Hz
modulation values are used to avoid interference between the signals
Remote command:
4.1.4.6
Setting the Measurement Time (MTime)
Access
: [MTime]
Access VNC
: [d]
The measurement time determines the interval at which new measurement results are
displayed. Internally, values are captured every 100
ms, that is: 10 per second. (With
the High Measurement Rate option R&S
EVSG-K22 installed, the internal capture rate
is 10
ms, that is: 100 per second.) If the defined measurement time is longer, the val-
ues captured internally in that interval are averaged and only the average value is dis-
played.
Note that the measurement time determines the minimum interval for values to be
stored during recording (see
Enter the time in milliseconds.
Remote command:
4.1.4.7
DDM and SDM Configuration
The following settings configure the DDM and SDM measurement results.
DDM Unit
Specifies the unit of the DDM display.
1 (dimensionless value) | % |
μ
A
Remote command:
SDM Unit
Specifies the unit of the SDM display.
1 (dimensionless value) | % |
μ
A
Remote command:
ILS Localizer and ILS Glidepath (GP) Modes