![Rohde & Schwarz EVSG-B1 User Manual Download Page 164](http://html.mh-extra.com/html/rohde-and-schwarz/evsg-b1/evsg-b1_user-manual_1477820164.webp)
Graphic Measurement Modes
R&S
®
EVSG1000
164
User Manual 1178.6227.02 ─ 06
The amplitude settings configure the y-axis of the spectrum. The following settings are
available for LF input.
Grid Y Max
Defines the maximum value of the y-axis.
Y % /Div
Defines the value range displayed in each division of the y-axis. The full y-axis range is
eight times this value.
Grid Unit
Access
: [CONFIG]
Access VNC
: [s]
Switches the unit of the y-axis between V and % for LF input (see
Range
Access
: [AMPT]
Access VNC
: [w]
Defines the full scale AF signal level. Define the correct level to avoid clipping and
overload.
Switches the range between 1
V and 5
V.
Remote command:
Coupling
Access
: [AMPT]
Access VNC
: [w]
Specifies the handling of the DC component of the AF signal.
For low IF signals, AC coupling is always used.
"AC"
The DC component of the AF signal is not forwarded.
"DC"
The DC component of the AF signal is also forwarded and analyzed.
Remote command:
DC Reference
Access
: [AMPT]
Access VNC
: [w]
AF Time Domain Mode (Option R&S
EVSG-K12)