Transducer Correction
Configures the level correction. The specified value is added to the measured power
levels to compensate for an inherent offset by the measurement setup, for example the
antenna.
Remote command:
4.2.3.3
Setting the Measurement Time (MTime)
Access
: [MTime]
Access VNC
: [d]
The measurement time determines the interval at which new measurement results are
displayed. Internally, values are captured every 100
ms, that is: 10 per second. (With
the High Measurement Rate option R&S
EVSG-K22 installed, the internal capture rate
is 10
ms, that is: 100 per second.) If the defined measurement time is longer, the val-
ues captured internally in that interval are averaged and only the average value is dis-
played.
Note that the measurement time determines the minimum interval for values to be
stored during recording (see
Enter the time in milliseconds.
Remote command:
4.3
VOR Mode (Option R&S
EVSG-K2)
Access
: [Mode] > "VOR"
Access VNC
: [m] > Down arrow key
If the VOR Mode option (R&S
EVSG-K2) is installed, signals from VOR systems can
be analyzed with the R&S
EVSG1000.
Remote command:
VOR Mode (Option R&S
EVSG-K2)