NimbleGen MS 200 Microarray Scanner Operator’s Manual v1.0
2.
Open the instrument’s front cover by pulling with both hands on both
sides of the white part so that the front cover disengages and swings
open to the right (Figure 40).
Figure 40: Opening the Front Cover
3.
If accessible, manually remove the slide or slide fragments from the
slide transport (Figure 41). Use vacuum to remove all slide fragments if
necessary.
Figure 41: Removing the Slide from the Slide Transport
4.
Close the instrument’s front cover (Figure 42). Gently push the front
cover into place, making sure to push it all the way into place. Two
bolts and a pin secure the cover.
Figure 42: Closing Front Cover