NimbleGen MS 200 Microarray Scanner Operator’s Manual v1.0
| 53
Figure 29: Area Definition Control for Specifying Multiple Scan Areas
5.
Position the mouse pointer inside the magenta feature area rectangle.
Drag both bounding rectangles
to the upper-left corner of the
microarray features.
6.
Drag the merged rectangles to the top-left segment of the printed area
to identify the area to scan. Adjust the size of the magenta rectangle by
dragging a corner handle.
7.
Use the white scan area rectangle to enclose the top (left) feature
segment on the array. This is used to set size and shape of all other scan
areas. Drag a corner handle or a side handle until that bounding
rectangle matches the desired feature size.
Summary of Contents for Nimblegen MS 200