NimbleGen MS 200 Microarray Scanner Operator’s Manual v1.0
| 69
1.
Remove the ILC slide stored inside the Microarray Scanner:
a.
If necessary, switch on the Microarray Scanner.
b.
If necessary, eject the slide magazine and remove the slide inserted
into slot 48.
c.
Start the instrument software.
d.
Click the
QC
tab and then
ILC
in the instrument software. The
Integrated Laser Check window opens (Figure 16).
e.
Click the
Remove ILC slide
button ( ) in the Integrated Laser
Check window.
f.
Close the instrument software and switch off the Microarray
Scanner
.
2.
Disinfect the instrument as described in the “Instrument Disinfection”
section on page 70.
Treat all parts of the instrument that come into contact with
biological samples or other hazardous material as potentially
infectious areas.
STOP
Thoroughly disinfect the instrument before it is removed from
the laboratory.
3.
Pack the instrument. Use the original packaging to ensure that no
damage occurs to the instrument.
STOP
Remove the slides from the magazine before you prepare the
instrument for relocation. If a slide remains in the instrument,
potential jamming or fragmentation of the slide could damage
the instrument.
If the original packaging is lost or damaged, contact Roche
NimbleGen for replacement packaging.
4.
Complete the Certificate of Decontamination (refer to page 71) and
affix to the outside of the package.
Summary of Contents for Nimblegen MS 200
Page 1: ......
Page 2: ......
Page 22: ...10 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Page 42: ...30 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Page 84: ...72 NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 Notes ...
Page 95: ...NimbleGen MS 200 Microarray Scanner Operator s Manual v1 0 83 ...
Page 103: ......