Chapter 5 To Trigger the Oscilloscope
5-10
DS2000E User’s Guide
Pulse Trigger
In pulse trigger mode, the oscilloscope will trigger when the pulse width of the input
signal satisfies the specified pulse width condition.
Trigger Type:
Press Type, rotate
to select "Pulse" and press down
. At this point, the trigger
setting information is displayed at the upper right corner of the screen. For example,
. The trigger type is pulse trigger; the trigger source is CH1; the trigger
level is 0.00 V.
Source Selection:
Press Source to select CH1, CH2, or EXT. For the details, please refer to the
introduction in "Trigger Source". The current trigger source is displayed at the
upper-right corner of the screen.
Note: Select channel with signal input as trigger source to obtain stable trigger.
Pulse Condition:
In this oscilloscope, positive pulse width is defined as the time difference between
the two crossing points of the trigger level and positive pulse; negative pulse width is
defined as the time difference between the two crossing points of the trigger level
and negative pulse, as shown in the figure below.
Figure 5-3 Positive Pulse Width/Negative Pulse Width
Press When to select the desired pulse condition.
: trigger when the positive pulse width of the input signal is greater than
Positive
pulse width
Trigger level
Trigger level
Negative
pulse width
A
B
A
B
Summary of Contents for DS2000E Series
Page 1: ...RIGOL User s Guide DS2000E Series Digital Oscilloscope Oct 2017 RIGOL TECHNOLOGIES INC...
Page 2: ......
Page 22: ......
Page 62: ......
Page 72: ......
Page 82: ......
Page 198: ......
Page 212: ......
Page 238: ......
Page 258: ......
Page 261: ...Chapter 14 Remote Control RIGOL DS2000E User s Guide 14 3 3 4 5 6...
Page 268: ......
Page 283: ...Index RIGOL DS2000E User s Guide 3 X Y 3 4 X Y mode 6 37 Y Cursor 6 29 Y T 3 4...