Chapter 5 To Trigger the Oscilloscope
RIGOL
DS2000E User’s Guide
5-47
the "Bit Fill", "Answer Error", "Check Error", "Format Error" or "Random Error".
Baud:
Press Baud to set the CAN baud to match the CAN bus signal and the values
available are 10 kb/s (default), 20 kb/s, 33.3 kb/s, 50 kb/s, 62.5 kb/s, 83.3 kb/s, 100
kb/s, 125 kb/s, 250 kb/s, 500 kb/s, 800 kb/s, 1 Mb/s or User. After "User" is selected,
press User and use
or the inner knob of the navigation knob to modify the baud
rate at a relatively smaller step or turn the outer knob of the navigation knob to
adjust the baud rate at a relatively larger step. The adjustable range is from 10 kb/s
to 1 Mb/s.
Sample Point:
Sample point is a point within a bit’s time. The oscilloscope samples the bit level at
this point. The sample position is represented by the percentage of "the time from
the start of the bit’s time to the sample point time" in the "bit’s time".
Press SamplePoint and use
to modify the parameter with a step of 1% and the
range is from 5% to 95%.
Figure 5-18 Sample Position
Trigger Mode:
Press Sweep to open the trigger mode list and select "Auto", "Normal" or "Single".
For the details, please refer to "Trigger Mode". The corresponding status light of
the current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameters (noise rejection) under this trigger type.
Trigger Level:
Use TRIGGER LEVEL to modify the level. For the details, please refer to the
introduction of "Trigger Level".
1Bit
60%
70%
80%
Summary of Contents for DS2000E Series
Page 1: ...RIGOL User s Guide DS2000E Series Digital Oscilloscope Oct 2017 RIGOL TECHNOLOGIES INC...
Page 2: ......
Page 22: ......
Page 62: ......
Page 72: ......
Page 82: ......
Page 198: ......
Page 212: ......
Page 238: ......
Page 258: ......
Page 261: ...Chapter 14 Remote Control RIGOL DS2000E User s Guide 14 3 3 4 5 6...
Page 268: ......
Page 283: ...Index RIGOL DS2000E User s Guide 3 X Y 3 4 X Y mode 6 37 Y Cursor 6 29 Y T 3 4...