Chapter 6 MATH and Measurements
RIGOL
DS2000E User’s Guide
6-37
If multiple measurement parameters are turned on later, you can use Meas.Para. to
switch among at most five measurement parameters that are turned on last and the
measurement parameter currently selected is high-lighted at the bottom of the
screen.
The following figure shows the auto measurement of the period of CH1.
X-Y Mode
X-Y mode is only available when the main time base is "X-Y" mode. In this mode, you
can adjust the cursor to measure X and Y values of the crossing points of the two
pairs of cursors.
Press MENU at the horizontal control are (HORIZONTAL) on the front panel and
then press Time Base to select "X-Y". Press Cursor
Mode
"X-Y" to enable the
X-Y mode cursor measurement function. You can also press Cursor or Mode
continuously to switch to "X-Y". The measurement results are displayed at the upper
left corner of the screen in the following mode.
CurAX: the X value at cursor A.
CurBX: the X value at cursor B.