
RIGOL
Chapter 5 To Trigger the Oscilloscope
5-18
DS2000E User’s Guide
: trigger on the falling edge of the input signal when the voltage level meets
the specified trigger level.
Idle Time:
Press Idle to set the idle time before the edge counting in Nth edge trigger. Use
or the inner knob of the navigation knob to adjust the idle time at a relatively smaller
step or turn the outer knob of the navigation knob to adjust the idle time at a
relatively larger step. The range available is from 16 ns to 4 s.
Edge Number:
Press Edge to set the value of "N" in Nth edge trigger. Use
or the inner knob of
the navigation knob to adjust the number of edges at a relatively smaller step or turn
the outer knob of the navigation knob to adjust the number of edges at a relatively
larger step. The range available is from 1 to 65535.
Trigger Mode:
Press Sweep and select "Auto", "Normal" or "Single". For the details, please refer to
Trigger Mode". The corresponding status light of the current trigger mode turns
on.
Trigger Setting:
Press Setting to set the trigger parameters (noise rejection) under this trigger type.
Trigger Level:
Use TRIGGER LEVEL to modify the level. For the details, please refer to the
introduction of "Trigger Level".
Summary of Contents for DS2000E Series
Page 1: ...RIGOL User s Guide DS2000E Series Digital Oscilloscope Oct 2017 RIGOL TECHNOLOGIES INC...
Page 2: ......
Page 22: ......
Page 62: ......
Page 72: ......
Page 82: ......
Page 198: ......
Page 212: ......
Page 238: ......
Page 258: ......
Page 261: ...Chapter 14 Remote Control RIGOL DS2000E User s Guide 14 3 3 4 5 6...
Page 268: ......
Page 283: ...Index RIGOL DS2000E User s Guide 3 X Y 3 4 X Y mode 6 37 Y Cursor 6 29 Y T 3 4...