Chapter 5 To Trigger the Oscilloscope
RIGOL
DS2000E User’s Guide
5-5
Trigger Holdoff
Trigger holdoff can be used to stably trigger the complex waveforms. Holdoff time is
the amount of time that the oscilloscope waits after a trigger before re-arming the
trigger module. The oscilloscope will not trigger even if the trigger condition is met
during the holdoff time and will only re-arm the trigger module after the holdoff time
expires. For example, to stably trigger the repeat pulse series as shown in the figure
below, the holdoff time should be set to a value greater than T1 and lower than T2.
Figure 5-2 Schematic Diagram of Trigger Holdoff
Press MENU (in the trigger control area (TRIGGER) on the front panel)
Setting
Holdoff; then, turn
or the inner knob of the navigation knob to modify the
holdoff time at a relatively smaller step or turn the outer knob of the navigation knob
to adjust the holdoff time (the default is 100 ns) at a relatively larger step until the
trigger becomes stable. The adjustable range is from 100 ns to 10 s.
Note: For DS2000E, the trigger types that support trigger holdoff include edge
trigger, pulse trigger, runt trigger, windows trigger, slope trigger, pattern trigger,
delay trigger, and duration trigger.
T2
Holdoff Time
T1
Trigger Positions
Summary of Contents for DS2000E Series
Page 1: ...RIGOL User s Guide DS2000E Series Digital Oscilloscope Oct 2017 RIGOL TECHNOLOGIES INC...
Page 2: ......
Page 22: ......
Page 62: ......
Page 72: ......
Page 82: ......
Page 198: ......
Page 212: ......
Page 238: ......
Page 258: ......
Page 261: ...Chapter 14 Remote Control RIGOL DS2000E User s Guide 14 3 3 4 5 6...
Page 268: ......
Page 283: ...Index RIGOL DS2000E User s Guide 3 X Y 3 4 X Y mode 6 37 Y Cursor 6 29 Y T 3 4...