
Chapter 5 To Trigger the Oscilloscope
RIGOL
DS2000E User’s Guide
5-27
Delay Trigger (Option)
In delay trigger, you need to set signal source A and signal source B. The oscilloscope
triggers when the time difference (
△
T) between the specified edges of source A
(Edge A) and source B (Edge B) meets the preset time limit, as shown in the figure
below.
Note:
Edge A and Edge B must be neighbouring edges.
T
Source A
Source B
Edge A=
Edge B=
Figure 5-9 Delay Trigger
Trigger Type:
Press Type to open the trigger type list. Rotate
and press down
to select
"Delay". At this point, the trigger setting information is displayed at the upper right
corner of the screen. For example,
. The trigger type is delay
trigger; the trigger source is CH1; the trigger level is -1.80 V.
Source A:
Press SourceA to select CH1 or CH2 as the trigger source of signal source A. For the
details, please refer to the introduction in "
Trigger Source". The signal source
currently selected is displayed in the upper-right corner of the screen.
Note:
Select channel with signal input as trigger source to obtain stable trigger.
Edge A:
Press EdgeA to select the trigger edge type of signal source A in delay trigger. It can
be set to the rising edge (
) or falling edge (
).
Summary of Contents for DS2000E Series
Page 1: ...RIGOL User s Guide DS2000E Series Digital Oscilloscope Oct 2017 RIGOL TECHNOLOGIES INC...
Page 2: ......
Page 22: ......
Page 62: ......
Page 72: ......
Page 82: ......
Page 198: ......
Page 212: ......
Page 238: ......
Page 258: ......
Page 261: ...Chapter 14 Remote Control RIGOL DS2000E User s Guide 14 3 3 4 5 6...
Page 268: ......
Page 283: ...Index RIGOL DS2000E User s Guide 3 X Y 3 4 X Y mode 6 37 Y Cursor 6 29 Y T 3 4...