Chapter 5 To Trigger the Oscilloscope
5-2
DS2000E User’s Guide
Trigger Source
Press MENU
Source in the trigger control area (TRIGGER) on the front panel to
select the desired trigger source. Signals input from CH1, CH2, or the [EXT TRIG]
connector, as well as the AC line can all be used as the trigger source.
Analog channel input:
Signals input from analog channels CH1 and CH2 can all be used as the trigger
source. No matter whether the channel selected is enabled, the channel can work
normally.
External trigger input:
Signals (such as external clock and signal of the circuit to be tested) input from the
[EXT TRIG] connector can be used as the trigger source. When this type of trigger
source is selected, you can set the trigger level within -4 V to +4 V by rotating
TRIGGER LEVEL.
When 2 analog channels are used for data acquisition, the trigger signal input from
the [EXT TRIG] connector can be used as the trigger source.
AC line:
The trigger signal is obtained from the AC power input of the oscilloscope. AC trigger
is usually used to measure signals relevant to the AC power frequency. For example,
stably trigger the waveform output from the transformer of a transformer substation;
it is mainly used in related measurements of the power industry.
Summary of Contents for DS2000E Series
Page 1: ...RIGOL User s Guide DS2000E Series Digital Oscilloscope Oct 2017 RIGOL TECHNOLOGIES INC...
Page 2: ......
Page 22: ......
Page 62: ......
Page 72: ......
Page 82: ......
Page 198: ......
Page 212: ......
Page 238: ......
Page 258: ......
Page 261: ...Chapter 14 Remote Control RIGOL DS2000E User s Guide 14 3 3 4 5 6...
Page 268: ......
Page 283: ...Index RIGOL DS2000E User s Guide 3 X Y 3 4 X Y mode 6 37 Y Cursor 6 29 Y T 3 4...