RIGOL
1-6
DS2000E User’s Guide
Function Inspection
1.
Press Storage
Default to restore the oscilloscope to its factory state.
2.
Connect the ground alligator clip of the probe to the "Ground Terminal" as
shown in the figure below.
3.
Connect one end of the probe to CH1 input terminal of the oscilloscope, and
connect the other end to the "Probe Compensation Signal Output Terminal".
Figure 1-6 Probe Compensation Signal Output Terminal/Ground Terminal
4.
Set the probe attenuation to 10X, and then press AUTO.
5.
Observe the waveform on the display. If the waveforms actually shown do not
match that in the figure below, please perform "Probe Compensation".
Figure 1-7 Probe Compensation Signal
6.
Use the same method to test the other channels.
WARNING
To avoid electric shock when using the probe, please make sure that the
insulated wire of the probe is in good condition. Do not touch the metallic
part of the probe when the probe is connected to high voltage source.
Tip
The probe compensation signal can only be used for probe compensation
adjustment and cannot be used for calibration.
Probe Compensation Signal Output Terminal
Ground Terminal
Summary of Contents for DS2000E Series
Page 1: ...RIGOL User s Guide DS2000E Series Digital Oscilloscope Oct 2017 RIGOL TECHNOLOGIES INC...
Page 2: ......
Page 22: ......
Page 62: ......
Page 72: ......
Page 82: ......
Page 198: ......
Page 212: ......
Page 238: ......
Page 258: ......
Page 261: ...Chapter 14 Remote Control RIGOL DS2000E User s Guide 14 3 3 4 5 6...
Page 268: ......
Page 283: ...Index RIGOL DS2000E User s Guide 3 X Y 3 4 X Y mode 6 37 Y Cursor 6 29 Y T 3 4...