RIGOL
9-2
DS2000E User’s Guide
To Enable Pass/Fail Test
Press Utility
Pass/Fail
Enable to select "ON". The area covered by blue on
the screen denotes the failure area and if the waveform measured at certain time
point during the measurement process passes through the failure area, it will be
treated as failed frame. Press Operate to select " " to start testing and select "
■
"
to stop testing.
Note: The pass/fail test function is not available in the following status.
In X-Y time base mode;
In ROLL time base mode;
In Y-T time base mode and the horizontal time base is set to 200ms/div or slower,
namely the instrument enters "Slow Sweep" mode;
When the waveform record is enabled;
None of the analog channels (CH1, CH2) is turned on.
After the pass/fail test is enabled, you can set the output type of the test results
according to "To Set Test Results Output Type". You can also select the signal
source, set the test mask range, create mask as well as save and recall the test mask.
For the details, refer to the introductions below.
To Select Source
Press Source to select the channel (CH1 or CH2) to be tested and only channels
enabled can be selected.
Mask Range
Users can define their desired test masks. If the waveform passes through the mask
range (namely the waveform enters the area covered by blue), it is judged as a failed
frame.
Press Range
Range to select "Screen" or "Cursor". The default is "Screen".
When "Screen" is selected, press X Mask and Y Mask respectively and rotate
to
Summary of Contents for DS2000E Series
Page 1: ...RIGOL User s Guide DS2000E Series Digital Oscilloscope Oct 2017 RIGOL TECHNOLOGIES INC...
Page 2: ......
Page 22: ......
Page 62: ......
Page 72: ......
Page 82: ......
Page 198: ......
Page 212: ......
Page 238: ......
Page 258: ......
Page 261: ...Chapter 14 Remote Control RIGOL DS2000E User s Guide 14 3 3 4 5 6...
Page 268: ......
Page 283: ...Index RIGOL DS2000E User s Guide 3 X Y 3 4 X Y mode 6 37 Y Cursor 6 29 Y T 3 4...