Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
The RCS-100 is compliance for each EMISSIONS test specified by the standard, e.g.
EMISSIONS class and group.
Emissions
Compliance
Electromagnetic environment-- guidance
RF emissions
CISPR 11
Group 1
The RCS-100 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The RCS-100 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity
The RCS-100 is compliance for each IMMUNITY test specified by the standard, e.g. IMMUNITY test
level.
Immunity test
IEC 60601-1-2
test level
Compliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
8 kV contact
±
2 kV,
±
4 kV,
±
8 kV,
±
15 kV air
±
8 kV contact
±
2 kV,
±
4 kV,
±
8 kV,
±
15 kV air
Radiated RF EM fields
IEC 61000-4-3
10V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
Electrical fast
transient/burst
IEC 61000-4-4
±
2 kV
100 kHz repetition frequency
±
2 kV
100 kHz repetition frequency
Surge
IEC 61000-4-5
±
0.5 kV,
±
1 kV line-to-line;;
±
0.5 kV,
±
1 kV and
±
2 kV line-to-
ground;;
±
0.5 kV,
±
1 kV line-to-line;;
±
0.5 kV,
±
1 kV, and
±
2 kV line-to-
ground;;
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3 V
0.15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0.15 MHz and 80 MHz
80 % AM at 1 kHz
3 V
0.15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0.15 MHz and 80 MHz
80 % AM at 1 kHz
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0 % U
T
: 0.5 cycle
a)
At 0
o
, 45
o
, 90
o
, 135
o
, 180
o
, 225
o
,
270
o
, and 315
o
.
0 % U
T
: 0.5 cycle
a)
At 0
o
, 45
o
, 90
o
, 135
o
, 180
o
, 225
o
, 270
o
,
and 315
o
.
0 % U
T
: 1 cycle
70 % U
T
: 25/30 cycles
b)
Single phase: at 0
o
0 % U
T
: 1 cycle
70 % U
T
: 25/30 cycles
b)
Single phase: at 0
o
0 % U
T
: 250/300 cycles
b)
0 % U
T
: 250/300 cycles
b)
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz
NOTE a) U
T
is the a.c. mains voltage prior to application of the test level;;
b) e.g. 25/30 means 25 periods at 50 Hz or 30 periods at 60 Hz.
23
Summary of Contents for RCS-100
Page 1: ...RCS 100 Gebrauchsanweisung...
Page 13: ...RCS 100 Instruction For Use...
Page 25: ...RCS 100 Instructions d utilisation...
Page 37: ...RCS 100 Instrucciones de uso...
Page 49: ...RCS 100...
Page 51: ...51 2 3 3 3 1 3 2 3 6 2600 3 3 4 Riester RCS 100 5 Riester 7 3...
Page 54: ...54 1 2 2800 3 5600 4 4500 5 7500 6 10 000 7 9000 8 6500 2 3 5 7 7 a 1...
Page 60: ...60 1 RCS 100 CISPR 11 1 RCS 100 CISPR 11 B RCS 100 IEC 61000 3 2 A IEC 61000 3 3...
Page 63: ...63 RCS 100 Istruzioni per l uso...
Page 78: ...78...
Page 79: ...79...