10.2 682 Fieldbus Configuration
The following procedure details how to configure the 682 for Fieldbus. Once the necessary 682 parameters are configured and
installation complete, communication with the SCT-2200 Fieldbus option (firmware version 1.25 or higher) is ready.
1. Navigate to the Communications menu in
Setup
mode (
Communications
displays.
2. Press
.
Serial
displays.
3. Press
.
RS-232 Port 1
displays.
4. Press
or
until
RS-485
displays.
5. Press
.
Trigger
displays.
6. Press
. The current setting for the RS-485 Trigger is highlighted.
7. Ensure
Fieldbus
is selected and press
.
NOTE: The Trigger parameter for the Serial Option Card Port would be set to Fieldbus when using a serial option card.
8. Press
twice.
Serial
displays.
9. Press
.
Fieldbus
displays.
10. Press
.
Network Protocol
displays.
11. Press
. The current set Network Protocol is highlighted. Select needed protocol and press
.
12. Press
.
Protocol Parameters
displays. Edit the protocol specific parameters if necessary.
13. Press
to return to
Weigh
mode.
If the Fieldbus option is not communicating the correct data, adjust the Byte/Word Swap parameter in the indicator.
See
for the Fieldbus menu structure and parameter descriptions.
682 Fieldbus Troubleshooting
After configuring the 682 parameters, cycle power to the module if it is wired to the 682.
The EtherNet/IP, Modbus TCP, PROFINET, DevicNet, and CANopen modules have three status LEDs:
• Green LED = Power — Green Power LED is on when there is power to the module
• Yellow LED = Status — Yellow Status LED is the RS-485 communication to the 682
If the module is not communicating, the yellow LED blinks rapidly for ~15 seconds then turns off for 1 second.
If the module is communicating with the 682, the yellow LED blinks rapidly and does not turn off after ~15 seconds.
• Red LED = Error — Red Error LED blinks until PLC is connected and communicating with the module, then turns off
The PROFIBUS module has four status LEDs that function differently than the above modules:
• Red LED = Power — Red Power LED stays on when there is power to the module
• Yellow LED = PROFIBUS — Yellow PROFIBUS LED stays on once module is connected to the PLC
If the module is not connected to the 682, the Yellow PROFIBUS LED blinks rapidly for ~15 seconds, then the Green
RX LED blinks 3 times, followed by the Yellow TX LED blinking 3 times. This cycle repeats until module connects.
• Green LED = RX — Green RX LED and Yellow TX LED blink alternately when module is communicating with the 682
• Yellow LED = TX — Yellow TX LED and Green RX LED blink alternately when module is communicating with the 682