682 Synergy Plus
54
Visit our website
5.0
Split Mode Configuration
The 682 supports multi-range and multi-interval scales of either two or three ranges or intervals. The full scale capacity is the
second range/interval when just Range 1 is set or the third range/interval when Range 1 and Range 2 are set.
Figure 5-1. Setup – Scale Format Menu – Split Mode
Table 5-1. Split Mode Options
Mode
Description
Multi-Range
Provides two or three ranges, each extending to the maximum capacity specified for each range; the scale display division changes
as the applied weight increases, but does not reset to lower display divisions until the scale returns to center of zero
Multi-Interval
Divides the scale capacity into two or three partial weighing intervals, each with different display divisions; the intervals extend to the
maximum capacity specified for each interval; the display divisions change with both increasing and decreasing loads
Off
Disables split mode
Table 5-2. Split Mode – Sub-Parameters
Parameter
Description
Range 1 Capacity
(MRMI)
Sets the capacity of Range 1 for the scale in primary units; this parameter must be set for
Range 2 Capacity
and
Range 2 Format
to display;
Enter value: 0.0-9999999.0,
0.0
(default)
Range 1 Format
Sets the Range 1 units format (decimal point and display divisions);
Settings:
8888881
(default), 8888882, 8888885, 8888810,
8888820, 8888850, 8888100, 8888200, 8888500, 88.88881, 88.88882, 88.88885, 888.8881, 888.8882, 888.8885, 8888.881,
8888.882, 8888.885, 88888.81, 88888.82, 88888.85, 888888.1, 888888.2, 888888.5
Range 2 Capacity
(MRMI)
Sets the capacity of Range 2 for the scale in primary units; only displays if
Range 1 Capacity
is set to a value greater than 0;
Enter value: 0.0-9999999.0,
0.0
(default)
Range 2 Format
Sets the Range 2 units format (decimal point and display divisions); only displays if
Range 1 Capacity
is set to a value greater than
0;
Settings:
8888881
(default), 8888882, 8888885, 8888810, 8888820, 8888850, 8888100, 8888200, 8888500, 88.88881,
88.88882, 88.88885, 888.8881, 888.8882, 888.8885, 8888.881, 8888.882, 8888.885, 88888.81, 88888.82, 88888.85,
888888.1, 888888.2, 888888.5
Primary
Secondary
Custom Unit 3
Setup
Tare
User
Scale Format
Configuration
Calibration
Format
Units
Split Mode
Range 1 Capacity
Range 1 Format
Split Mode
Range 2 Capacity
Range 2 Format
Displays if
Split Mode
is set to
Multi-Range
or
Multi-Interval
Displays if
Range 1 Capacity
is set to a value greater than
0