Configuration
© Rice Lake Weighing Systems
●
All Rights Reserved
35
4.5.3 Calibration Menu
This section provides a flow chart and descriptions for the Calibration menu.
Figure 4-9. Calibration Menu
Table 4-8. Calibration Menu Parameter
Parameter
Description
Zero Calibration
Executes the zero calibration process (
)
Zero Counts
Displays the raw count value at the zero weight; A zero calibration generates this raw count value; Manually changing this count value
changes the zero weight and negates the zero calibration
Test Weight
Sets the test weight value for the span calibration (
);
Enter value: 0.000001–9999999.999999,
10000.0
(default)
Span Calibration Executes the span calibration process (
)
Span Counts
Displays the raw count value at the span weight; A span calibration generates this raw count value; Manually changing this count value
changes the span weight and negates the span calibration
Linear Calibration Linear or multi-point calibration is performed by entering up to four additional calibration points (
);
Points 1-4 have the following sub-parameters:
Linear Point # Weight – Sets the test weight value for the linear calibration point
Calibrate Linear Point # – Executes the linear calibration process for the point; generates the raw count value for the test weight
value for Point #
Linear Point # Counts – Displays the raw count value at the linear point weight; A linear calibration generates this raw count
value; Manually changing this count value changes the linear point weight and negates the linear calibration for the point
Rezero
Removes an offset value from the zero and span calibrations (
)
Last Zero
Takes last push button zero in the system (from
Weigh
mode) and uses it as the new zero reference point, after which a new span
calibration must be performed; this calibration cannot be performed when calibrating a scale for the first time (
)
Temp Zero
Temporarily zeros the displayed weight of a non-empty scale, after a span calibration was performed;
The difference between the temporary zero and the previously calibrated zero value is used as an offset (
)
Point 3
Point 4
Point 2
Point 1
Linear Calibration
Rezero
Temp Zero
Point 4
Last Zero
Zero Calibration
Zero Counts
Span Calibration
Test Weight
Span Counts
Temp Zero
Setup
Tare
User
Calibration
Scale Format
Communications