Configuration
There are two types of configuration parameters in the 682,
Setup
mode parameters (or Legal for Trade configuration) and
User
mode parameters (or non-legal configuration).
Setup
mode parameters are accessed by pressing the setup switch (
).
User
mode parameters are accessed by pressing the menu button and do not require pressing the setup switch. Some
User
mode parameters can be password protected (
The following sections provide graphic representations of the 682 menu structures. Most menu diagrams are accompanied by a
table which describes all parameters and parameter values associated with the menu. The factory default setting appears at the
top of each column in bold type.
The Audit, Calibration, Setpoints, Accumulator, Tare, Time, Date, MAC ID, Fieldbus Version and Version menus can be
accessed by pressing the MENU button. The Audit, Tare, MAC ID, Fieldbus Version and Version menus are read only and
accessible in the top-level menu. The Calibration, Setpoints, Accumulator, Time and Date menus are available in the User
menu and can be password protected (
).
NOTE: The User mode setpoints menu displays the setpoint value of configured setpoints and is accessible with the menu
button. Complete configuration of setpoints is available in Setup mode within the Setup menu.
The Setup menu is accessed by pressing the setup switch (
) and can also be password protected.
NOTE: All weight related parameters must be configured prior to calibrating the unit.
4.1 Setup Switch
In order to configure the 682, it must be placed in
Setup
mode with the setup switch or with the audit jumper (
).
The setup switch is accessed through a small hole on the bottom of the enclosure. Remove the setup switch screw and insert a
non-conductive tool into the access hole to press the setup switch.
IMPORTANT: Use caution when inserting the non-conductive tool into the enclosure. Insert the tool about 3/4 in (19 mm),
until the switch is engaged. Do not use excessive force which could damage the switch.
Figure 4-1. Setup Switch Access
When the 682 is placed in
Setup
mode, the Setup menu is accessed and
Configuration
for a detailed breakdown of this menu. Torque the setup switch screw to 10 in-lb (1.1 N-m) when reinserting.
4.1.1 Audit Jumper
The audit jumper (J24) turns
Setup
mode access ON and OFF. Access to
Setup
mode is allowed without pressing the setup
switch when the audit jumper is in the ON position. Access to
Setup
mode requires pressing the setup switch when the audit
jumper is in the OFF position. See
for the location of the audit jumper on the CPU board.
NOTE: In certain Legal for Trade applications it is necessary to seal the indicator to restrict access to the setup switch
). Breaking of the seal terminates the Legal for Trade status of the indicator.
Setup Switch
Access