6.6 Calibration Jumper
Calibration access can be restricted by removing the calibration shunt.
Figure 6-11. Restricting Calibration Access
6.7 Messages and Errors
6.7.1 Programming Errors
Message
Description
...
If the WWP radios are type-approved, the display shows the platform number instead of the weight
Weight reset in progress; contact technical support if the message persists
Unstable weight when acquiring a point during the calibration phase
(flashing)
Weighing error; unload the platform, turn it off and turn on the indicator again; contact technical support if the problem persists
(flashing)
If the weight exceeds the maximum capacity, immediately remove the load and check that the platform has not been damaged
Message
Description
Solution
Calibration error
First calibrate the zero point, then proceed with the next points
Calibration error
Check the connection of the load cell
Check that the cell signal is stable, valid and greater than the previously acquired point
Calibration error
Increase the calibration weight
Calibration error
Check that the signal coming from the cell increases as the weight on the scale increases
When acquiring the calibration points, use the increasing calibration weights
Calibration error
Check that the signal coming from the load cell is not negative
Calibration error
Repeat the calibration, checking that the capacity and division have been correctly set
Calibration error
Capacity entered in the
parameter is not a multiple of the division set in the
step
Instrument not configured
Reset factory configurations (menu
, parameter
,
Section 4.1.2 on page 31
)
Calibration error
Maximum capacity not set (
=
)
Calibration error
Reset factory configurations (menu
, parameter
,
Section 4.1.2 on page 31
)
Instrument configured but not calibrated
Perform calibration
Weight unstable
Check
in the
menu (
Table 4-1 on page 30
) that the signal is stable and
retry
if connection of the cells is with 4 wires, check that the sense jumpers are inserted
2
5
3
1
4