Load Ranger Wheel Weigh Pad System
38
Visit our website
www.RiceLake.com
5.0
Calibration
This section provides the procedures necessary to calibrate the Load Ranger wheel weigh pads through the built-in displays
and with the Ai-1 indicator.
NOTE: When using the Ai-1 indicator with the weigh pads, skip to
Section 5.2 on page 40
to calibrate all of the weigh pads
through the Ai-1 indicator.
5.1 Calibrate with Wheel Weigh Pad Display
This section describes procedures for setting calibration parameters and the calibration of the Load Ranger Wheel Weigh Pads
through the built-in display.
5.1.1 Complete Calibration Procedure
NOTE: Press
at any time to back out one level or multiple times to return to the Weigh mode.
1. Turn indicator on by pressing and press during startup.
displays.
2. Press .
displays.
NOTE: If
displays instead of
, press
, then press
two times until
displays. Press
and
displays then continue with
Step 3
.
3. Press .
displays.
4. Press . The current decimal setting displays.
5. Press to scroll through decimal settings and press to save selection.
displays.
6. Press . The current division setting displays.
7. Press to scroll through division settings and press to save selection.
displays.
8. Press . The current unit setting displays.
9. Press to scroll through unit settings and press to save selection.
displays.
10. Press . Range 1 value displays. Use the numeric entry procedure to set value.
NOTE: For single-range wheel weigh pads, Range 1 is set as the max wheel weigh pad capacity.
11. Press .
displays.
NOTE: Range 2 should only be set for multi-range wheel weigh pads. For single-range wheel weigh pads, skip to
Step 14
.
12. Press . Range 2 value displays. Use the numeric entry procedure to set value.
13. Press .
displays.
14. Press .
displays.
15. Press .
displays. Ensure wheel weigh pad is unloaded.
16. Press .
briefly displays once the procedure is complete, then
displays.
17. Press . The current setting for the number of calibration points displays.
18. Press to scroll through settings (1-3) and press to save selection. Calibration point 1 value displays.
19. Use the numeric entry procedure to set the calibration weight value and press .
displays.