Remote Control Commands
R&S
®
NRX
386
User Manual 1178.5566.02 ─ 07
MEMory:TABLe:GAIN[:MAGNitude]
<value>...
Defines the offset values in the selected correction table. Existing data is overwritten.
Take care that the number of frequency values matches the number of offset values,
defined by
. If the numbers differ, excess values are
ignored.
Use
to select the table.
Setting parameters:
<value>
Numeric values with a maximum of 3
digits after the decimal
point, separated by commas. Values with more than 3
decimal
places are rounded.
Default unit: dB
Example:
MEM:TABL:GAIN 0,0.0033,0.04
Sets 3 offset values; 0.000
dB, 0.003
dB, 0.040
dB.
Usage:
Setting only
Manual operation:
See
MEMory:TABLe:MAP
<register_name>[, <register_number>]
Assigns a name to the correction table stored in the memory location.
Take care to use recognizable names. The R&S NRX does not check whether a name
is unique. If you want to reset the factory default, use
.
The query returns the memory location that is assigned to the setup name.
Parameters:
<register_name>
Name of the correction table; allowed are alphanumeric charac-
ters and special characters.
<register_number>
Memory location of the correction table
0
Factory-set correction table, cannot be changed
1 to 9
Available memory locations
Example:
MEM:TABL:MAP "test5",5
Example:
MEM:TABL:MAP? "test5"
Manual operation:
See
MEMory:TABLe:MOVE
<string>...
Renames of the selected correction table.
Use
to select the table.
Managing Setups and Correction Tables