Sensor Configuration
R&S
®
NRX
131
User Manual 1178.5566.02 ─ 07
Video Bandwidth
For measuring the peak envelope power, specify the video bandwidth that the power
sensor uses for measuring the detected RF signal.
"4
kHz" | "200
kHz" | "Full"
"Full" means that the maximum bandwidth of the power sensor is
used.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:VBWidth[:VALue]
[SENSe<Sensor>:]BANDwidth:VIDeo:FNUMber
[SENSe<Sensor>:]BWIDth:VIDeo:FNUMber
Integration Time Mode
Specifies which integration time is used for a single measurement.
"Default"
Uses the default settings.
"User"
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:APERture:MODE
Integration Time
Defines the integration time for a single measurement.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:APERture[:VALue]
Averaging Mode
Sets the averaging mode.
"User"
Define the value under
.
"Auto"
Determines the average count automatically from the level of the
input signal.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO[:
STATe]
Averaging Count
is set to "User".
Sets the number of readings that are averaged for one measured value. The higher the
count, the lower the noise, and the longer it takes to obtain a measured value.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt[:VALue]
NRT Measurement Type
www.allice.de
Allice Messtechnik GmbH