System Settings
R&S
®
NRX
147
User Manual 1178.5566.02 ─ 07
"Trigger Out"
Available for the Out
1
/
Trig
Out BNC connector "I/O
1" tab.
Provides a trigger signal at the Out
1
/
Trig
Out connector. Select the
trigger source under
Trigger Source for Trigger Output
.
"Trigger In"
Available for the Trig
In
/
Out
2 BNC connector ("I/O
2" tab).
Apply an external trigger signal at the Trig
In
/
Out
2 connector. Set
the termination resistance under
.
Remote command:
[SENSe<Sensor>:]POWer:REFLection:RANGe:LIMit[:STATe]
[SENSe<Sensor>:]POWer[:POWer]:RANGe:LIMit[:STATe]
Measurement for Recorder Output ← I/O 1, I/O 2 tabs
Available if
is set to:
●
"Analog Out"
●
"Forw Analog Out"
●
"Refl Analog Out"
Sets the measurement of which the results are output.
Remote command:
OUTPut:RECorder<output>:FEED:INDex
0 V Equivalent ← I/O 1, I/O 2 tabs
Available if
is set to:
●
"Analog Out"
●
"Forw Analog Out"
●
"Refl Analog Out"
Enter the measurement value that corresponds to 0
V output voltage.
Remote command:
OUTPut:RECorder<output>:LIMit:LOWer:CCDF
OUTPut:RECorder<output>:LIMit:LOWer:POWer
OUTPut:RECorder<output>:LIMit:LOWer:RATio:RCOefficient
OUTPut:RECorder<output>:LIMit:LOWer:RATio:RFRatio
OUTPut:RECorder<output>:LIMit:LOWer:RATio:RLOSs
OUTPut:RECorder<output>:LIMit:LOWer:RATio:SWR
OUTPut:RECorder<output>:LIMit:LOWer:RATio[:VALue]
OUTPut:RECorder<output>:LIMit:LOWer[:VALue]
2.5 V Equivalent ← I/O 1, I/O 2 tabs
Available if
is set to:
●
"Analog Out"
●
"Forw Analog Out"
●
"Refl Analog Out"
Enter the measurement value that corresponds to 2.5
V output voltage.
Remote command:
OUTPut:RECorder<output>:LIMit:UPPer:CCDF
OUTPut:RECorder<output>:LIMit:UPPer:POWer
OUTPut:RECorder<output>:LIMit:UPPer:RATio:RCOefficient
OUTPut:RECorder<output>:LIMit:UPPer:RATio:RFRatio
Connections