Remote Control Commands
R&S
®
NRX
397
User Manual 1178.5566.02 ─ 07
[SENSe<Sensor>:]ADD
<sensor>
Adds a LAN power sensor. See
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Setting parameters:
<sensor>
Hostname of the power sensor.
Example:
ADD "NRQ6-101435"
Usage:
Setting only
Manual operation:
See
SYSTem:SENSor<Sensor>:INFO?
[<argument>]
Queries information about the selected power sensor, including calibration data.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Query parameters:
<argument>
System Information and Configuration
www.allice.de
Allice Messtechnik GmbH