Measurement Types and Result Displays
R&S
®
NRX
93
User Manual 1178.5566.02 ─ 07
Start of Gate
Sets the start time of the gate.
Remote command:
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG]:TIME
Length of Gate
Sets the length of the gate.
Remote command:
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG]:OFFSet[:TIME]
Fence
Enables or disables an exclusion interval for the selected gate. This exclusion interval
is called fence. The interval where the fence overlaps with the gate is excluded from
the measurement.
Like a gate, a fence is defined by its start time and its length.
Remote command:
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG][:EXCLude]:MID[:
STATe]
Start of Fence
Sets the start time of the fence. The start time refers to the start of the gate.
Remote command:
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG][:EXCLude]:MID:
TIME
Length of Fence
Sets the length of the fence.
Remote command:
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG][:EXCLude]:MID:
OFFSet[:TIME]
7.6
Timeslot
Supported by multipath and wideband power sensors. The power sensor measures the
average power in successive timeslots. You can define the number and characteristics
of the timeslots. But different to time gate measurements, where each gate is defined
individually, the timeslots all share the characteristics. Thus, the timeslots are arranged
in a frame structure with equal spacing and are suitable for periodic signals. The time
Timeslot
www.allice.de
Allice Messtechnik GmbH