Remote Control Commands
R&S
®
NRX
432
User Manual 1178.5566.02 ─ 07
<Gate>
1 to 4
Time gate
Parameters:
<time_interval>
Default unit: s
[SENSe<Sensor>:][POWer:]TGATe<Gate>[:EXCLude]:MID:TIME
<time_interval>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG][:EXCLude]:MID:
TIME
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
<Gate>
1 to 4
Time gate
Parameters:
<time_interval>
Default unit: s
[SENSe<Sensor>:][POWer:]TGATe[:EXCLude]:MID[:STATe]
<state>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG][:EXCLude]:MID[:
STATe]
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:COUNt
<count>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:POWer]:TSLot[:AVG]:COUNt
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<count>
Range:
1 to 128
*RST:
8
R&S NRP2 Compatibility
www.allice.de
Allice Messtechnik GmbH