Remote Control Basics
R&S
®
NRX
481
User Manual 1178.5566.02 ─ 07
14.4.8.8
Operation Upper Limit Fail Status Registers
The
CONDition
registers contain information whether a measured value currently
exceeds a configured upper limit. The
EVENt
registers contain information whether a
measured value exceeded an upper limit value since the last query.
For information on index
Chapter 14.4.1, "Hierarchy of the Status Registers"
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Extension Summary
Measurement m+1 Upper Limit Fail
Measurement m+2 Upper Limit Fail
Measurement m+3 Upper Limit Fail
Measurement m+4 Upper Limit Fail
Measurement m+5 Upper Limit Fail
Measurement m+6 Upper Limit Fail
Measurement m+7 Upper Limit Fail
Measurement m+8 Upper Limit Fail
0
0
0
0
0
0
0
+
Figure 14-17: Operation upper limit fail status registers
Querying the register:
●
STATus:OPERation:ULFail<RegisterIndex>:CONDition?
●
STATus:OPERation:ULFail<RegisterIndex>[:SUMMary][:EVENt]?
Table 14-19: Used operation upper limit fail status bits and their meaning
Bit no.
Short description
Bit is set if measured values of a measurement channel
exceed the upper limit value or for the summary.
0
Extension summary
Sum bit of the next lower status register.
1
Measurement m+1 upper limit fail
Channel
1
2
Measurement m+2 upper limit fail
Channel
2
3
Measurement m+3 upper limit fail
Channel
3
4
Measurement m+4 upper limit fail
Channel
4
5
Measurement m+5 upper limit fail
Channel
5
6
Measurement m+6 upper limit fail
Channel
6
Status Reporting System
www.allice.de
Allice Messtechnik GmbH