Remote Control Commands
R&S
®
NRX
273
User Manual 1178.5566.02 ─ 07
[SENSe<Sensor>:]STATistics:SCALe:X:MPWidth?
Queries the greatest level resolution that can be attained. If the value is exceeded, a
settings conflict occurs, due to the following reasons:
●
Number of pixels that has been selected is too great, see
CALCulate<Measurement>:STATistics[:SCALe]:X:POINts
●
Width chosen for the level range is too small, see
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Query only
CALCulate<Measurement>:STATistics:PDF[:SCALe]:Y:PDIVision
<value>
CALCulate<Measurement>:STATistics[:SCALe]:Y:SPACing
LIN
is set.
Sets the scaling of the y-axis for PDF statistics measurements.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
0.01 to 1000.0
*RST:
0.2
Default unit: -
Manual operation:
See
CALCulate<Measurement>:STATistics:PDF[:SCALe]:Y:TOP
<value>
CALCulate<Measurement>:STATistics[:SCALe]:Y:SPACing
LIN
is set.
Sets the maximum value of the y-axis for PDF statistics measurements.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
0.0 to 10000.0
*RST:
1.0
Default unit: -
Manual operation:
See
Measurement Settings and Results
www.allice.de
Allice Messtechnik GmbH