Remote Control Commands
R&S
®
NRX
275
User Manual 1178.5566.02 ─ 07
CALCulate<Measurement>:STATistics[:SCALe]:X:POINts
<value>
Effective for statistics measurements.
Sets the measurement result resolution. It specifies the number of pixels that are
assigned to the logarithmic level range for measured value output. The width of the
level range divided by N–1, where N is the number of pixels, must not be less than the
smallest level resolution.
●
Level range:
CALCulate<Measurement>:STATistics[:SCALe]:X:RANGe
●
[SENSe<Sensor>:]STATistics:SCALe:X:
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
3 to 8191
*RST:
600
Manual operation:
CALCulate<Measurement>:STATistics[:SCALe]:X:RANGe
<value>
Effective for statistics measurements.
Sets the width of the level range for the analysis result.
Lower limit of the level range:
●
CALCulate<Measurement>:STATistics[:SCALe]:X:RLEVel:RELative
●
CALCulate<Measurement>:STATistics[:SCALe]:X:RLEVel[:ABSolute]
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
0.01 to 100.0
*RST:
50.0
Default unit: dB
Manual operation:
CALCulate<Measurement>:STATistics[:SCALe]:X:RLEVel:RELative
<value>
Effective for statistics measurements with relative power display.
Measurement Settings and Results