Remote Control Commands
R&S
®
NRX
385
User Manual 1178.5566.02 ─ 07
MEMory:TABLe:DATA:POINts?
Queries the number of data pairs (frequency - offset) in the selected table.
to select the table.
Usage:
Query only
MEMory:TABLe:FREQuency
<value>...
Defines the frequency values in the selected correction table. Existing data is overwrit-
ten. Take care that the number of frequency values matches the number of offset val-
ues, defined by
. If the numbers differ, excess
values are ignored.
to select the table.
Setting parameters:
<value>
Numeric values with a maximum of 2
digits after the decimal
point, separated by commas. Values with more than 2
decimal
places are rounded.
Default unit: Hz
Example:
MEM:TABL:FREQ 50.00,60,70.3456
Sets 3 frequency values; 50.00
Hz, 60.00
Hz, 70.35
Hz.
Usage:
Setting only
Manual operation:
MEMory:TABLe:FREQuency:POINts?
Queries the number of frequency values in the selected correction table.
to select the table.
Usage:
Query only
Manual operation:
MEMory:TABLe:GAIN:POINts?
Queries the number of offset values in the selected correction table.
to select the table.
Usage:
Query only
Manual operation:
Managing Setups and Correction Tables