Remote Commands
R&S
®
FSMR3000
441
User Manual 1179.0116.02 ─ 02
*OPC should be used after requesting data. This will hold off any subsequent changes
to the selected trigger source, until after the sweep is completed and the data is
returned.
Configuring the Triggering Conditions
...................................................................441
Configuring the Trigger Output
..............................................................................444
12.5.6.1
Configuring the Triggering Conditions
The following commands are required to configure a triggered measurement.
...........................................................................................441
TRIGger[:SEQuence]:HOLDoff[:TIME]
..............................................................................441
TRIGger[:SEQuence]:IFPower:HOLDoff
........................................................................... 441
TRIGger[:SEQuence]:IFPower:HYSTeresis
.......................................................................442
TRIGger[:SEQuence]:LEVel[:EXTernal<port>]
...................................................................442
TRIGger[:SEQuence]:LEVel:IFPower
............................................................................... 443
TRIGger[:SEQuence]:LEVel:IQPower
...............................................................................443
TRIGger[:SEQuence]:LEVel:RFPower
..............................................................................443
.......................................................................................... 443
........................................................................................444
TRIGger[:SEQuence]:DTIMe
<DropoutTime>
Defines the time the input signal must stay below the trigger level before a trigger is
detected again.
Parameters:
<DropoutTime>
Dropout time of the trigger.
Range:
0 s to 10.0 s
*RST:
0 s
Default unit: S
Manual operation:
See
TRIGger[:SEQuence]:HOLDoff[:TIME]
<Offset>
Defines the time offset between the trigger event and the start of the measurement.
Parameters:
<Offset>
*RST:
0 s
Default unit: S
Example:
TRIG:HOLD 500us
Manual operation:
See
TRIGger[:SEQuence]:IFPower:HOLDoff
<Period>
This command defines the holding time before the next trigger event.
Configuring the Measurement