Remote Commands
R&S
®
FSMR3000
641
User Manual 1179.0116.02 ─ 02
*RST:
0
Example:
INST:COUP:USER2:STAT ON
Turns on the coupling with the index number 2.
Manual operation:
See
INSTrument:COUPle:USER<uc>:WINDow:LIST?
[<ChannelName>, <Parameter>]
This command queries the measurement windows that can be synchronized with
another channel (or measurement window).
Note that synchronizing with a specific measurement window is only possible in the
Analog Demodulation application.
Suffix:
<uc>
.
irrelevant
Query parameters:
<ChannelName>
Optional SCPI parameter.
String containing the name of a measurement channel.
<Parameter>
Optional SCPI parameter.
String containing the name of a parameter that you can syn-
chronize.
Return values:
<SynchronizableWindow>
<char_data>
'All Windows'
All windows can be synchronized. This value is always returned
if no parameters are provided with the command.
Comma-separated list of strings
String containing the names of the measurement windows that
can be synchronized.
This value is only available for marker coupling, which can be
set independently of the measurement window.
Example:
INST:COUP:USER:WIND:LIST?
Result:
'All Windows'
Example:
INST:COUP:USER:WIND:LIST? 'Analog
Demod','Frequency Marker 1'
Result:
'All Windows','1','2','3','4','5','6'
Usage:
Query only
Manual operation:
See
Configuring the R&S
FSMR3