Remote Commands
R&S
®
FSMR3000
516
User Manual 1179.0116.02 ─ 02
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:LEFT 10MHz
Sets the left limit of the search range to 10 MHz.
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt
<SearchLimit>
This command defines the right limit of the marker search range for
all
markers in
all
windows.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:RIGH 20MHz
Sets the right limit of the search range to 20 MHz.
CALCulate<n>:MARKer<m>:X:SLIMits:ZOOM[:STATe]
<State>
This command adjusts the marker search range to the zoom area for
all
markers in
all
windows.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK:X:SLIM:ZOOM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:RIGH 20MHz
Sets the right limit of the search range to 20 MHz.
CALCulate<n>:THReshold
<Level>
This command defines a threshold level for the marker peak search (for
all
markers in
all
windows).
Note that you must enable the use of the threshold using
Suffix:
<n>
.
irrelevant
Analyzing Measurements (Basics)