Common Measurement Settings
R&S
®
FSMR3000
154
User Manual 1179.0116.02 ─ 02
Remote command:
OUTPut<up>:TRIGger<tp>:PULSe:IMMediate
7.5.2
Demod Output
Access
: "Overview"
≥
"Output" > Demod Output
The demodulated signal in time domain results can be output to the IF/VIDEO/DEMOD
output connector on the R&S
FSMR3000.
Note that the audio frequency (AF) filter settings used for demodulation also apply to
the online output. However, a maximum of two high pass, low pass or deemphasis fil-
ters can be active at the same time if demod output is active.
Online Demodulation Output State
............................................................................. 154
Online Demodulation Output State
Enables or disables online demodulation output. If enabled, the demodulated audio fre-
quencies are output to the IF/VIDEO/DEMOD output connector on the rear panel of the
R&S
FSMR3000.
Remote command:
OUTPut<up>:ADEMod[:ONLine][:STATe]
Output Selection
Selects the result display whose results are output. Only time domain results can be
selected. All currently active time domain result displays are listed.
Output