Getting Started
R&S
®
FSMR3000
40
User Manual 1179.0116.02 ─ 02
Figure 4-6: Calibration signal with optimized display settings
4.4.2
Displaying a Spectrogram
In addition to the standard "level versus frequency" spectrum display, the R&S
FSMR3
also provides a spectrogram display of the measured data. A spectrogram shows how
the spectral density of a signal varies over time. The x-axis shows the frequency, the y-
axis shows the time. A third dimension, the power level, is indicated by different colors.
Thus you can see how the strength of the signal varies over time for different frequen-
cies.
1. Tap the "Overview" softkey to display the general configuration dialog box.
2. Tap the "Display Config" button.
The SmartGrid mode is activated, and the evaluation bar with the available evalua-
tion methods is displayed.
3.
Drag the "Spectrogram" icon from the evaluation bar to the diagram area. The blue
area indicates that the new diagram would replace the previous spectrum display.
Since we do not want to replace the spectrum, drag the icon to the lower half of the
display to add an additional window instead.
Trying Out the Instrument