Remote Commands
R&S
®
FSMR3000
502
User Manual 1179.0116.02 ─ 02
Parameters:
<Trace>
Number of the trace to be stored
(This parameter is ignored for
ALL
).
<FileName>
String containing the path and name of the import file.
MMEMory:STORe<n>:TRACe
<Trace>, <FileName>
This command exports trace data from the specified window to an ASCII file.
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
Suffix:
<n>
.
Parameters:
<Trace>
Number of the trace to be stored
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR1:TRAC 1,'C:\TEST.ASC'
Stores trace 1 from window 1 in the file
TEST.ASC
.
Manual operation:
See
MMEMory:STORe<n>:IQ:STATe
<Number>, <FileName>
This command stores the currently captured I/Q data to a file.
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
Suffix:
<n>
.
1..n
irrelevant
Setting parameters:
<Number>
Always '1'.
<FileName>
String containing the path and file name.
The file type is .iq.tar.
Analyzing Measurements (Basics)