Remote Commands
R&S
®
FSMR3000
517
User Manual 1179.0116.02 ─ 02
Parameters:
<Level>
Numeric value. The value range and unit are variable.
*RST:
-120 dBm
Default unit: DBM
Example:
CALC:THR:STAT ON
Example:
CALC:THR -82DBM
Enables the search threshold and sets the threshold value to -82
dBm.
CALCulate<n>:THReshold:STATe
<State>
This command turns a threshold for the marker peak search on and off (for
all
markers
in
all
windows).
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:THR:STAT ON
Switches on the threshold line.
12.6.4.4
Positioning the Marker
This chapter contains remote commands necessary to position the marker on a trace.
..................................................................................517
..................................................................................... 519
Positioning Normal Markers
The following commands position markers on the trace.
CALCulate<n>:MARKer<m>:MAXimum:LEFT
...................................................................517
CALCulate<n>:MARKer<m>:MAXimum:NEXT
.................................................................. 518
CALCulate<n>:MARKer<m>:MAXimum[:PEAK]
................................................................ 518
CALCulate<n>:MARKer<m>:MAXimum:RIGHt
..................................................................518
CALCulate<n>:MARKer<m>:MINimum:LEFT
....................................................................518
CALCulate<n>:MARKer<m>:MINimum:NEXT
................................................................... 519
CALCulate<n>:MARKer<m>:MINimum[:PEAK]
................................................................. 519
CALCulate<n>:MARKer<m>:MINimum:RIGHt
...................................................................519
CALCulate<n>:MARKer<m>:MAXimum:LEFT
This command moves a marker to the next lower peak.
Analyzing Measurements (Basics)