Remote Commands
R&S
®
FSMR3000
442
User Manual 1179.0116.02 ─ 02
Note that this command can be used for
any trigger source
, not just IF Power
(despite the legacy keyword).
Parameters:
<Period>
Range:
0 s to 10 s
*RST:
0 s
Default unit: S
Example:
TRIG:SOUR EXT
Sets an external trigger source.
TRIG:IFP:HOLD 200 ns
Sets the holding time to 200 ns.
Manual operation:
See
TRIGger[:SEQuence]:IFPower:HYSTeresis
<Hysteresis>
This command defines the trigger hysteresis, which is only available for "IF Power" trig-
ger sources.
Parameters:
<Hysteresis>
Range:
3 dB to 50 dB
*RST:
3 dB
Default unit: DB
Example:
TRIG:SOUR IFP
Sets the IF power trigger source.
TRIG:IFP:HYST 10DB
Sets the hysteresis limit value.
Manual operation:
See
TRIGger[:SEQuence]:LEVel[:EXTernal<port>]
<TriggerLevel>
This command defines the level the external signal must exceed to cause a trigger
event.
Suffix:
<port>
.
Selects the trigger port.
1 = trigger port 1(TRIGGER INPUT/OUTPUT connector on front
panel)
2 = trigger port 2 (TRIGGER INPUT/OUTPUT connector on rear
panel)
Parameters:
<TriggerLevel>
Range:
0.5 V to 3.5 V
*RST:
1.4 V
Default unit: V
Example:
TRIG:LEV 2V
Manual operation:
See
Configuring the Measurement