Remote Commands
R&S
®
FSMR3000
443
User Manual 1179.0116.02 ─ 02
TRIGger[:SEQuence]:LEVel:IFPower
<TriggerLevel>
This command defines the power level at the third intermediate frequency that must be
exceeded to cause a trigger event.
Note that any RF attenuation or preamplification is considered when the trigger level is
analyzed. If defined, a reference level offset is also considered.
Parameters:
<TriggerLevel>
For details on available trigger levels and trigger bandwidths see
the data sheet.
*RST:
-20 dBm
Default unit: DBM
Example:
TRIG:LEV:IFP -30DBM
TRIGger[:SEQuence]:LEVel:IQPower
<TriggerLevel>
This command defines the magnitude the I/Q data must exceed to cause a trigger
event.
Note that any RF attenuation or preamplification is considered when the trigger level is
analyzed. If defined, a reference level offset is also considered.
Parameters:
<TriggerLevel>
Range:
-130 dBm to 30 dBm
*RST:
-20 dBm
Default unit: DBM
Example:
TRIG:LEV:IQP -30DBM
TRIGger[:SEQuence]:LEVel:RFPower
<TriggerLevel>
This command defines the power level the RF input must exceed to cause a trigger
event. Note that any RF attenuation or preamplification is considered when the trigger
level is analyzed. If defined, a reference level offset is also considered.
The input signal must be between 500
MHz and 8
GHz.
Parameters:
<TriggerLevel>
For details on available trigger levels and trigger bandwidths see
the data sheet.
*RST:
-20 dBm
Default unit: DBM
Example:
TRIG:LEV:RFP -30dBm
TRIGger[:SEQuence]:SLOPe
<Type>
Parameters:
<Type>
POSitive | NEGative
Configuring the Measurement