Rotor-Gene Q MDx CE User Manual 02/2022
39
In the event that updates are required due to uncovered security vulnerabilities, QIAGEN provides
mechanisms to install a defined set of validated Windows security patches either online (if internet
connection is available on the QIAGEN laptop) or as offline package, prepared on a separate
computer with internet connection.
Please visit the product page on
QIAGEN.com
for more information.
4.2
Site requirements
Rotor-Gene Q MDx instruments must be located out of direct sunlight, away from heat sources, and
away from sources of vibration and electrical interference. Refer to Appendix A for the operating
conditions (temperature and humidity). The installation site should be free of excessive drafts,
excessive moisture, excessive dust, and not subject to large temperature fluctuations.
Refer to Appendix A for the weight and dimensions of Rotor-Gene Q MDx instruments. Ensure that
the workbench is dry, clean, and has additional space for accessories. For further information about
required specifications of the workbench, contact QIAGEN Technical Services.
Note
: It is extremely important that the Rotor-Gene Q MDx instrument is placed on a stable surface,
which is level and vibration free. Refer to operating conditions — see Appendix A.
The Rotor-Gene Q MDx instrument must be placed within approximately 1.5 m (59 in.) of a properly
grounded (earthed) AC power outlet.
WARNING
Explosive atmosphere
The Rotor-Gene Q MDx instrument is not designed for use in an explosive
atmosphere.
WARNING
Risk of overheating
To ensure proper ventilation, maintain a minimum clearance of 10 cm at the
rear of the Rotor-Gene Q MDx instrument.
Slits and openings that ensure the ventilation of the Rotor-Gene Q MDx
instrument must not be covered.