Rotor-Gene Q MDx CE User Manual 02/2022
162
o
Product Sheet
Rotor-Gene Series Software Version 1.7 or above, which contains the easy-to-use OTV Rotor
wizard
Rotor-Disc 72 Rotor
Rotor-Disc 72 Locking Ring
9.3
Running an OTV
1.
Place the fluorescent insert over the emission lens in the bottom of the Rotor-Gene Q MDx
chamber.
2.
Place the OTV Rotor-Disc into a Rotor-Disc 72 Rotor. Secure using a Rotor-Disc 72 Locking
Ring. Place the assembly into the Rotor-Gene Q MDx and click into place. Close the Rotor-
Gene Q MDx lid.
3.
Access the Advanced wizard by selecting the
Advanced
tab in the
New Run
window. In the
Advanced wizard, click on
Instrument maintenance
and then
OTV
. The wizard prompts for the
OTV serial number, this is found on the OTV ring. Then click
Start
.