Propaq CS Directions for Use
81
De
fi
brillator
Sync
hr
onization
Warning
If the R-wave synchronization markers do not appear to be nearly simultaneous with the R-waves on the
Propaq CS monitor display or are not present, do not proceed with synchronized cardioversion.
You must press the LIFEPAK 5 Defibrillator SYNC button and check for appropriate synchronization markers
on the Propaq CS monitor before each attempt at cardioversion. Protocol Systems cannot guarantee the delay from
the sync marker to the defibrillator discharge.
Note
A problem with the Interface Cable connecting the defibrillator and monitor, such as a cable fault or unplugging the
cable, will prevent display of synchronization markers and may prevent the defibrillator from entering the synchronized mode.
5.
Follow hospital procedures and LIFEPAK 5 Defibrillator instructions for cardioversion.
6.
If subsequent cardioversion must be performed, repeat steps 3 through 5.
Remove the Interface Cable
1.
Disconnect the Interface Cable from the monitor.
2.
Press the lever on the side of the LIFEPAK 5
Defibrillator and slide the Interface forward until
it is detached.
3.
Store the Interface Cable in its static-protected
plastic bag.
Summary of Contents for Propaq 242
Page 6: ...6 Protocol Systems Inc...
Page 18: ...18 Protocol Systems Inc...
Page 26: ...26 Protocol Systems Inc...
Page 50: ...50 Protocol Systems Inc...
Page 52: ...52 Protocol Systems Inc...
Page 64: ...64 Protocol Systems Inc...
Page 66: ...66 Protocol Systems Inc...
Page 72: ...72 Protocol Systems Inc...
Page 74: ...74 Protocol Systems Inc...
Page 78: ...78 Protocol Systems Inc...
Page 84: ...84 Protocol Systems Inc...
Page 86: ...86 Protocol Systems Inc...
Page 94: ...94 Protocol Systems Inc...
Page 120: ...120 Protocol Systems Inc...