54
Protocol Systems, Inc.
Customize Alarm Limits Based on Patient’s Current Vital Signs
1.
To quickly set all alarm limits, from the Main Menu press
SETUP
,
ALARMS, STAT SET
. The
monitor turns on all alarms and calculates new alarm limits based on the patient’s current
vital sign values. Make sure that the new limits are appropriate for the patient.
Warning
If a patient’s vital sign value falls outside of the upper or lower alarm range limit, STAT SET turns off the alarm
and the alarm limit except for the following:
1. The lower alarm limits for SpO
2
and ETCO
2
are not turned off by STAT SET.
2. If HR/PR ALARM LIMITS in the Settings window is set to CANNOT TURN OFF, STAT SET affects HR/PR alarm limits
as follows:
HR/PR PATIENT VALUE
DISPLAY
UPPER LIMIT
LOWER LIMIT
Overrange
+++
Maximum
Unchanged
Underrange
- - -
Unchanged
Minimum
Indeterminate
???
Unchanged
Unchanged
STAT SET Limit Calculations
1
1. New alarm limits calculated by STAT SET cannot be outside the allowable alarm limit range. If a new limit is calculated
to be above or below the allowable alarm limit range, it defaults to the maximum or minimum alarm limit allowed for that
vital sign.
Vital Sign
If the Patient’s Vital Sign
Value is
Then Calculated New
Lower Limit is
Then Calculated New Upper
Limit is
Heart Rate
HR
≤
99
100 - 250
HR
≥
251
HR x 0.8
HR - 20
Unchanged
HR x 1.2
HR + 20
250
Pulse Rate
PR
≤
99
PR
≥
100
PR x 0.8
PR - 20
PR x 1.2
PR + 20
Invasive Pressure
Inv Prs
≤
25
26 - 99
Inv Prs
≥
100
Inv. Pressure - 5
Inv. Pressure x 0.8
Inv. Pressure - 20
Inv. Pr 5
Inv. Pressure x 1.2
Inv. Pr 20
NIBP
NIBP
≤
25
26 - 99
NIBP
≥
100
NIBP - 5
NIBP x 0.8
NIBP - 20
NIBP + 5
NIBP x 1.2
NIBP + 20
Respiration
Rate/Breath Rate
RR/BR
≤
25
26 - 99
RR/BR
≥
100
RR/BR - 5
RR/BR x 0.8
RR/BR - 20
RR/BR + 5
RR/BR x 1.2
RR/BR + 20
Temperature
Temp
≥
0˚C
Temp - 0.5
Temp + 0.5
SpO
2
SpO
2
≥
0%
SpO
2
- 5
(min. limit 50%)
100% (adult and pediatric mode)
SpO
2
+ 5 (neonate mode)
ETCO
2
ETCO
2
≥
0 mmHg
ETCO
2
≥
2.0 (% or kPa)
ETCO
2
- 5 mmHg
(min. 15 mmHg)
ETCO
2
- 0.7 (% or kPa)
(min 2.0% or 2.0 kPa)
ETCO
2
+ 10 mmHg
ETCO
2
+ 1.4 (% or kPa)
INCO
2
INCO
2
≥
0 mmHg
INCO
2
≥
0 (% or kPa)
Not affected by STAT SET
INCO
2
+ 5 mmHg
INCO
2
+ 0.7 (% or kPa)
Apnea Delay
Not affected by STAT SET
Summary of Contents for Propaq 242
Page 6: ...6 Protocol Systems Inc...
Page 18: ...18 Protocol Systems Inc...
Page 26: ...26 Protocol Systems Inc...
Page 50: ...50 Protocol Systems Inc...
Page 52: ...52 Protocol Systems Inc...
Page 64: ...64 Protocol Systems Inc...
Page 66: ...66 Protocol Systems Inc...
Page 72: ...72 Protocol Systems Inc...
Page 74: ...74 Protocol Systems Inc...
Page 78: ...78 Protocol Systems Inc...
Page 84: ...84 Protocol Systems Inc...
Page 86: ...86 Protocol Systems Inc...
Page 94: ...94 Protocol Systems Inc...
Page 120: ...120 Protocol Systems Inc...